  • ModelYSF9
  • BrandYSENMED
  • PriceGet Price
  • MOQ1
  • Ship FromFrom Guangzhou, China
  • ShippingAny shipping terms convenient for customers
  • -
    +
    Categories:Patient Monitors
    Share:
    Description Get Price Review


    Features

    Core
    · Newly advanced A9 main board with Linux OS, calculating speed is 4 times faster than traditional products
     
    · Support storage of 2160 hours trend table and graph review, 2 hours waveform review,
    2000 groups NIBP review and 2000 alarm events review
     
    Body
    · 15 inch high-brightness TFT LED
    · Support display 9~13 waveforms
    · Support 7 channel ECG waveform display simultaneously  
    · Optional HDMI output
     
    Printer
    · Built-in high-speed 50mm thermal printer
     (Brand: SEIKO, Japan)
     
    Central System
    · Optional Build-in wireless network module, supporting wired or wireless connection to the central monitoring station.
    · Optional Support HL7 (Health Level Seven)
     
     
    Alarm
    · Double high light alarm light bar.
    · Three-level acousto-optic alarm
    · Sensor-off alarm
    · Paper out alarm
    · Support alarm review
    · Support alarm pause
     
    Linux OS
    · Support operation with mouse and computer keyboard (option)
     
    · Multi-display mode to meet different clinical requirement, including standard interface, big font, OxyCRG, trend graph, NIBP review and full leads ECG interface.
     
    · NIBP self-test mode: including overpressure test, static pressure test and air leakage test.
    · Generate ID automatically when register a new patient
    · Support medical history search by patient ID, name and mobile number
    · 18 types of Arrhythmia analysis and real-time S-T segment analysis and pacemaker detection
    · Drug calculation and titration table 
    · Support multi-language display
    · Support online software upgrading by net/USB

    Interface
    Transducer socket
    · ECG
    · SPO2
    · NIBP  
    · IBP 1 / IBP 2
    · TEMP 1
    · TEMP 2 / Mainstream EtCO2 / AG (multi-gas)
    · Sidestream EtCO2
     
    LAN / Central monitor system interface
    · The LAN port is for connecting the monitor to a central monitor system network.
     
    Input device interface
    · This optional interface provides an USB ports to enable the monitor to be connected to off-the-shelf input devices:
    · Mouse: any specified trackball or USB mouse may be used for navigation and data entry.
    · Computer keyboard: an USB computer keyboard can be used for data entry instead of the on-screen pop-up keyboard.
    [Remark: user has to restart the machine after plug and unplug the mouse / computer keyboard.]
     
    · Knob (Standard)
     
     
    Service Features
    · The Support Tool helps technical personnel to
    –  carry out configuration, upgrades and troubleshooting on an individual monitor.
    – back up the monitor settings.
     
    · The Service Mode is password-protected and ensures that only trained staff can access service tests and tasks. It includes Skin Type, Brightness Setting, Clear Data and NIBP Calibration.
    · The Support Tool uses the USB interface of the monitor for software upgrading.

    Performance Specifications
     
    Dimension and Weight
    - Dimension: 393mm*382mm*173mm
    - Weight: 4.5kg(excluding accessories)
     
    Power Supply
    - Voltage: AC100~240V, 50/60HZ, Power≤60W
     
    Display
    - 15" color TFT LED  
    - Resolution: 1024*768 pixels
     
    Battery (Pluggable)
    - Type: Rechargeable lithium battery 14.8V/2200mAh
    - Charge Cycle: ≥500 times
    - Working time: 2 hours (optional larger battery for 4-5 hours)
     
    Recorder (Option)
    - Method: Thermal printer
    - Paper width: 50 mm (1.97 in)
    - Printing speed: 12.5/ 25/ 50 mm/s
    - Trace: Max. 3 tracks
    - Recording way: Real-time Recording, Review Printing,  
       Periodic Recording, Alarm Recording
     
    Alarm
    - Level: Low, medium and high
    - Indication: Auditory and visual
    - Alarm volume adjustable
    - Alarm pause time: 1min, 2min
    - Parameter alarm type: Latch/ Unlatch
     
    Input Device
    - Knob (Standard)
    - Keypad input (Standard)
    - Mouse/ Keyboard input (option)
     
    System Output & Extensible Interface
    - Ethernet Network: 1 standard RJ45 socket
    - USB Port: 1
    - Video Output:  1 HDMI port (option)
     
    Operating Environment
    - Temperature:  5 ~ 40 ℃
    - Humidity:  15% ~ 80% (non-condensing)
    - Atmosphere pressure:  86 KPa ~ 110 KPa
     
     
     
     
    Transportation and Storage Environment
    - Temperature:  -20~50 ℃
    - Humidity:  10%~90%  (non-condensing)
    - Atmosphere pressure:  86 KPa ~ 110 KPa
     
    Safety
    - IEC60601-1 Approved, CE marking according to MDD93/42/EEC
    - With reference to RoHS Directive 2011/65/EU recasting
     
    Trend & Reviewing
    - Trend: 2160 hours
    - ARR events: 128 groups of ARR events and associated waveform
    - NIBP measurement reviewing: 2000 groups
    - Waveform review: 2 hours
    - Alarm event: 2000 groups of parameter alarms events and associated parameter
     
    ECG
    - Lead mode: 3/5 Leads, I, II, III or I, II, III, AVR, AVL, AVF, V
    - Protection: Breakdown Voltage 4000VAC 50/60Hz; Defibrillator proof
    - Gain: 2.5mm/mV (×0.25), 5.0mm/mV (×0.5), 10mm/mV (×1), 20mm/mV (×2), 40 mm/ mV (×4), Auto
    - Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
    - ECG signal range:  ±12 mV (Gain × 0.25)
    - Accuracy: ±1bpm/ ±1%, whichever is greater
    - Resolution: 1 bpm
    - Leakage Current < 10 μA
    - Baseline Recovery:
    ≤ 3s after defibrillation (Monitor mode)
    ≤ 1s after defibrillation (Surgery mode)
    - Bandwidth:Surgery 1 ~ 25 Hz
    Monitor 0.5 ~ 40 Hz
    Diagnostic 0.05 ~ 120 Hz
    - Indication of Electrode Separation: Every electrode (exclusive of RL)
     
    Heart Rate
    - Measure range:
    Adult: 0bpm; 10 ~ 300 bpm
    Neo/Ped: 0bpm; 10 ~ 350 bpm
    - Resolution: 1 bpm
    - Accuracy: ± 1% or ± 1bpm, whichever is greater
     
     
     
     
    ST Measurement
    - Range: -2.0 ~ +2.0 mV
    - Accuracy: -0.8mV~+0.8mV: ±0.02mV or ±10%, whichever is greater
    - Other range: unspecified
    - Resolution: 0.01mV
     
    Respiration
    - Method: Impedance between RA-LL, RA-LA
    - Gain: ×0.25, ×0.50, ×1, ×2, ×4
    - Respiration Rate: 0bpm, 6 ~ 150 BrPM
    - Sweep speed:  6.25 mm/s, 12.5 mm/s, 25mm/s
    - Resolution: 1 BrPM
    - Accuracy: ±2 bpm or ±2%, whichever is greater
    - Apnea Alarm: 10 ~ 60 s
     
    NIBP
    - Method: Oscillometric
    - Measure mode: Manual, Auto, STAT
    - Measure Interval in AUTO Mode 1~480 min
    - STAT mode cycle time: Keep 5 minutes, at 5 seconds interval
    - Measure and Alarm Range:
    Adult: SYS:  30 ~ 270 mmHg
              DIA:  10 ~ 220 mmHg     
              MEAN:  20 ~ 235 mmHg
    Pediatric: SYS:  40 ~ 235 mmHg
    DIA:  10 ~ 220 mmHg     
    MEAN:  20 ~ 225 mmHg
    Neonate:  SYS:  30 ~ 135 mmHg
    DIA:  10 ~ 110 mmHg     
    MEAN:  20 ~ 125 mmHg
    - Static pressure accuracy: ±3mmHg
    - Resolution: 1mmHg
    - Accuracy: Maximum Mean error ±5mmHg           
             Maximum Standard deviation ≤8mmHg
    - Over pressure Protection: Dual protection via software & hardware
     
    Temperature
    - Technique: Thermistor probe (2.25K)
    - Channel: Dual-Channel, provide T1; T2; ΔT
    - Measuring and Alarm Range:
    0.0 ℃~ 50 ℃ (32℉~ 122℉)
    - Unit: Celsius (℃), Fahrenheit ( ℉)
    - Resolution: 0.1℃or 1℉
    - Accuracy: ±0.1℃(exclusive probe)

    Performance Specifications
     
    SpO2 (Digital Technic)
    - Measurement Range: 0 ~ 100 %
    - Resolution: 1 %
    - Response Modes: Low, Medium, High
    - Accuracy:  ±2% (70% ~ 100%)
                ±3 % (35% ~ 69%)
    Unspecified (0 ~ 34%)
    - Support Pitch tone and multi-level volume   
     
    Pulse Rate
    - Measuring and Alarm Range: 25~250bpm
    - Accuracy: ±1% or ±1 bpm, whichever is greater
    - Resolution: 1bpm
     
    Nellcor-SpO2 (Option)
    - Measurement Range: 0 ~ 100 %
    - Resolution: 1 %
    - Accuracy:   70% ~ 100%, ±2 % (adult)
                 70% ~ 100%, ±3 % (Neonate)
                 70% ~ 100%, ±2 % (Low Perfusion)
                 0% ~ 69%, unspecified
     
    Pulse Rate
    - Measurement range: 20 ~ 300 bpm
    - Resolution: 1bpm
    - Accuracy:
    ±3 bpm (20 ~ 250 bpm)
    unspecified (251~300 bpm)
     
    Masimo SpO2 (Option)
    - - Measurement Range: 0 ~ 100 %
    - - Resolution: 1 %
    - Accuracy:  
    70% ~ 100%, ±2 % (adult/ pediatric, non-motion conditions)
    70% ~ 100%, ±3 % (neonate, non-motion conditions)
    70% ~ 100%, ±3 % (motion conditions)
    0% ~ 69% unspecified
    Pulse Rate
    - Measurement range: 25 ~ 240 bpm
    - Resolution: 1bpm
    - Accuracy: ±3 bpm (non-motion condition)
     
     
     
     
     
     
     
    EtCO2 (Mainstream/ Sidestream) (Option)
    - Measure method:  Non-dispersive infrared (NDIR)
    - Measure Range:  0 ~19.7% (0 ~ 150 mmHg)
                     0 ~ 20 kPa
    - Resolution: 0.1 mmHg
    - CO2 Accuracy:  
    0 ~ 40 mmHg, ±2 mmHg
    41 ~ 70 mmHg, ±5% of reading
    71 ~ 100 mmHg, ±8% of reading
    101~ 150 mmHg, ±10% of reading
    at 760 mmHg, ambient temperature of 25℃)
    - Respiratory Rate:  Range: 3 ~150 BrPM
                    Accuracy: ±1 bpm
     
    EtCO2 (Micro-stream) (Option)
    - Measure method:  Non-dispersive infrared (NDIR)
    - Measure Range:  0 ~19.7% (0 ~150 mmHg)
                     0 ~ 20 kPa
    - Sample Rate: 50 mL/min ±10mL/min
    - Resolution:  0.1 mmHg (0 ~ 50 mmHg)
               0.25 mmHg (50 ~ 114 mmHg)
    - CO2 Accuracy:  0 ~ 40 mmHg, ±2 mmHg
                    41 ~ 70 mmHg, ±5% of reading
                    71 ~ 100 mmHg, ±8% of reading
                   101~ 150 mmHg, ±10% of reading
            at 760 mmHg, ambient temperature of 35℃)
    - Respiratory Rate: Range: 3 ~120 BrPM
                  Accuracy: ±1 bpm
     
    Depth of Anesthesia (CSI) (Option)
    - EEG sensitivity:  ± 400μV
    - Noise: < 2μVp-p, < 0.4μV RMS, 1-250 Hz
    - CMRR: > 140 dB
    - Input impedance: > 50 Mohm
    - Sample rate: 2000 samples/sec, (14 bits equivalent)
    - BS%: 0-100, filter 1-42 Hz, 1 sec. display update
    - EMG: 0-100 Logarithmic. Filter 75-85 Hz, 1 sec. update
    - Alarms: High / Low with user selectable limit
    - Artifact rejection: Automatic
    - Sensor impedance range: 0 - 10 kOhm / measurement current 0.01 μA
     

     
    Multi-gas/O2 (Anesthetic Gas) (Option)
    - Method: Infrared absorption
    - Gas sorts: CO2, N2O, Des, Iso, Enf, Sevo, Hal,
       O2 (Optional paramagnetic sensor)
    - Calibration: Room air calibration performed automatically when changing airway
    Airway adapt (< 5 sec)
    - Measurement range:
    CO2: 0~25%,  N2O : 0~100%                    
    O2: 0~100%,  Enf, Iso, Hal: 0~25%,
    Sevo, Des: 0~25%
    - Data output: Fi and ET values
    - Respiration rate: 0~ 150 BrPM
    - Other: Up to 5 waveforms displayed
          Agent mixture detection
      MAC value displayed
     
    IBP (Option)
    - Max Channel: 2
    - Measurement way: Thermal resistance way
    - Press Sensor:  Sensitivity   5 uV/V/mmHg, ±2%
                  Impedance   300 to 3000Ω
    - Resolution: 1 mmHg
    - Unit: mmKg, kPa, cmH2O
    - Transducer sites:  
    Arterial Pressure (ART)
    Pulmonary Arterial (PA)
    Left Arterial (LAP)
    Right Arterial (RAP)
    Central Venous Pressure (CVP)
    Intracranial Pressure (ICP)
    P1/ P2
    - Measuring and alarm range:
    ART   0 ~ +350mmHg
    PA    -10 ~ +120 mmHg
    CVP/ RAP/ LAP/ ICP-10 ~ +40 mmHg
    P1/ P2-50 ~ +350mmHg
    - Accuracy:  
    Static: ±1mmHg or ±2%, whichever is greater (exclusive of transducer)
          ±4mmHg or ±4%, whichever is greater (inclusive of transducer)
    Dynamic:  ±4mmHg or ±4%, whichever is greater
     
    The YSF9 patient monitor is designed to match the pace and unique needs of adult, pediatric,
    and neonatal intensive care; anesthesia and peri-operative care; OR and cardiac care environments.
     
    An optional USB Mouse/Keyboard control eases navigation and data entry.
     
    An optional HDMI output eases observation from a long distance.
    Portable Biphasic External Defibrillator YS-8000D
    Portable Biphasic External Defibrillator YS-8000D
    Write down your demand below and we will contact you within 24h!

    1/3
    X