We use cookies to improve your online experience. By continuing browsing this website, we assume you agree our use of cookies.
Categories

Medical hospital equipment Multi-parameter patient monitor YSF3

Model: YSF3
Medical hospital equipment Multi-parameter patient monitor YSF3
  • ModelYSF3
  • BrandYSENMED
  • PriceGet Price
  • MOQ1
  • Ship FromFrom Guangzhou, China
  • ShippingAny shipping terms convenient for customers
  • -
    +
    Categories:Patient Monitors
    Share:
    Description Get Price Review

     

    The  YSF3 vital signs monitor is engineered for the mobile, precise and fast recording of vital signs
    data at clinics, hospitals and outpatient treatments. Ergonomical design and highly developed technology
    resulting in reliable measurement results and time efficiency.
    12V DC power supply for ambulance use
     

    An optional HDMI output eases observation from a long distance.
     
    Core
    Newly advanced A8 main board with Linux OS, calculating speed is 4 times faster than traditional products
     
    Support storage of 2160 hours trend table and graph review, 2 hours waveform review, 2000 groups
    NIBP review and 2000 alarm events review
     
    Body
    7 inch high-brightness TFT LED
    Standard touch screen configuration
    Optional HDMI output
     
     
    Printer
    USB External high-speed 50mm thermal printer
     (Brand: SEIKO, Japan) 
     
    Central System
    Optional Build-in wireless network module, supporting wired or wireless connection to the central monitoring station. 
    Optional Support HL7 (Health Level Seven)
     
    Alarm
    Three-level acousto-optic alarm 
    Sensor-off alarm 
    Paper out alarm 
    Support alarm review 
    Support alarm pause
    Linux OS
    Multi-display mode to meet different clinical requirement
     
    NIBP self-test mode: including overpressure test, static pressure test and air leakage test.
     
    Generate ID automatically when register a new patient
    Support medical history search by patient ID, name and mobile number 
    Optional infrared thermometry allows clinicians to obtain patient’s temperature using non-invasive infrared technology that detects heat from the skin’s surface
     
    18 types of Arrhythmia analysis and real-time S-T segment analysis and pacemaker detection
     
    Drug calculation and titration table 
     
    Support multi-language display 
     
    Support online software upgrading by net/USB
     
    Suitable for monitoring adult, pediatric and neonatal patient. 
     
     
    Interface
    Transducer socket
    ECG 
    SPO2 
    NIBP  
    TEMP 1 
    Mainstream EtCO2 
     
    LAN / Central monitor system interface
    The LAN port is for connecting the monitor to a central monitor system network. 
     
    Input device interface 
    This optional interface provides an USB ports to enable the monitor to be connected to off-the-shelf input devices:
     
    Mouse: any specified trackball or USB mouse may be used for navigation and data entry.
    Computer keyboard: an USB computer keyboard can be used for data entry instead of the on-screen pop-up keyboard.
    [Remark: user has to restart the machine after plug and unplug the mouse / computer keyboard.]
     
    Touch screen (Standard)
    [Remark: Touch screen can not be used simultaneously with mouse / computer keyboard.]
     
    Service Features
    The Support Tool helps technical personnel to
    –  carry out configuration, upgrades and troubleshooting on an individual monitor.
    –back up the monitor settings.
     
    The Service Mode is password-protected and ensures that only trained staff can access service tests and tasks. It includes Skin Type, Brightness Setting, Clear Data, NIBP Calibration and Touch Screen Calibration.
    The Support Tool uses the USB interface of the monitor for software upgrading.
    Related Product
     
    Infrared thermometry
     
     
    Suntech NIBP module
     
     
    Respironics EtCO2 module
     
     
    Nellcor SpO2 module
     
     
    Masimo Spo2 module
     
    Performance Specifications
     
    Dimension and Weight
    -Dimension: 210mm*130mm*75mm
    -Weight: 1kg(excluding accessories)
     
    Power Supply
    -Voltage: AC100~240V, 50/60HZ
    -Adapter: Output 9VDC, 5A
     
    Display
    -7" color TFT LED  
    -Resolution: 800*480 pixels
     
    Battery  
    -Type: Rechargeable lithium battery 7.4V/2200mAh
    -Charge Cycle: ≥500 times
    -Working time: 2-3 hours (optional larger battery for 4-5 hours)
     
    Recorder (Option)
    -Method: USB External Thermal printer
    -Paper width: 50 mm (1.97 in)
    -Printing speed: 12.5/ 25/ 50 mm/s
    -Trace: Max. 3 tracks
    -Recording way: Real-time Recording, Review Printing,  
        Periodic Recording, Alarm Recording
     
    Alarm
    -Level: Low, medium and high
    -Indication: Auditory and visual
    -Alarm volume adjustable
    -Alarm pause time: 1min, 2min
    -Parameter alarm type: Latch/ Unlatch
     
    Input Device
    -Touch screen (standard)
    -Touch pen (standard)
     
    System Output & Extensible Interface
    -Ethernet Network: 1 standard RJ45 socket
    -USB Port: 1
    -Video Output: 1 HDMI port (option)
     
    Operating Environment
    -Temperature:  5 ~ 40 ℃
    -Humidity:  15% ~ 80% (non-condensing)
    -Atmosphere pressure:  86 KPa ~ 110 KPa
     
     
    Transportation and Storage Environment
    -Temperature:  -20~50 ℃
    -Humidity:  10%~90%  (non-condensing)
    -Atmosphere pressure:  86 KPa ~ 110 KPa
     
    Safety
    -IEC60601-1 Approved, CE marking according to MDD93/42/EEC
    -With reference to RoHS Directive 2011/65/EU recasting
     
    Trend & Reviewing
    -Trend: 2160 hours
    -ARR events: 128 groups of ARR events and associated waveform
    -NIBP measurement reviewing: 2000 groups
    -Waveform review: 2 hours
    -Alarm event: 2000 groups of parameter alarms events and associated parameter 
     
    ECG
    -Lead mode: 3/5 Leads, I, II, III or I, II, III, AVR, AVL, AVF, V
    -Protection: Breakdown Voltage 4000VAC 50/60Hz; Defibrillator proof
    -Gain: 2.5mm/mV (×0.25), 5.0mm/mV (×0.5), 10mm/mV (×1), 20mm/mV (×2), 40 mm/ mV (×4), Auto 
    -Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
    -ECG signal range:  ±12 mV (Gain × 0.25)
    -Accuracy: ±1bpm/ ±1%, whichever is greater
    -Resolution: 1 bpm
    -Leakage Current < 10 μA
    -Baseline Recovery: 
    ≤ 3s after defibrillation (Monitor mode)
    ≤ 1s after defibrillation (Surgery mode)
    -Bandwidth: Surgery 1 ~ 25 Hz
    Monitor 0.5 ~ 40 Hz
    Diagnostic 0.05 ~ 120 Hz
    -Indication of Electrode Separation: Every electrode (exclusive of RL)
     
    Heart Rate
    -Measure range: 
    Adult: 0bpm; 10 ~ 300 bpm
    Neo/Ped: 0bpm; 10 ~ 350 bpm
    -Resolution: 1 bpm
    -Accuracy: ± 1% or ± 1bpm, whichever is greater
     
     
     
    ST Measurement
    -Range: -2.0 ~ +2.0 mV
    -Accuracy: -0.8mV~+0.8mV: ±0.02mV or ±10%, whichever is greater
    -Other range: unspecified
    -Resolution: 0.01mV
     
    Respiration
    -Method: Impedance between RA-LL, RA-LA
    -Gain: ×0.25, ×0.50, ×1, ×2, ×4
    -Respiration Rate: 0bpm, 6 ~ 150 BrPM
    -Sweep speed:  6.25 mm/s, 12.5 mm/s, 25mm/s
    -Resolution: 1 BrPM
    -Accuracy: ±2 bpm or ±2%, whichever is greater
    -Apnea Alarm: 10 ~ 60 s
     
    Temperature
    -Technique: Thermistor probe (2.25K)
    -Channel:  Single-channel 
    -Measuring and Alarm 
    Range: 0.0 ℃~ 50 ℃ (32℉~ 122℉) 
    -Unit: Celsius (℃), Fahrenheit ( ℉)
    -Resolution: 0.1℃or 1℉
    -Accuracy: ±0.1℃(exclusive probe)
     
    NIBP
    -Method: Oscillometric
    -Measure mode: Manual, Auto, STAT
    -Measure Interval in AUTO Mode 1~480 min
    -STAT mode cycle time: Keep 5 minutes, at 5 seconds interval
    -Measure and Alarm Range: 
    Adult:  SYS:  30 ~ 270 mmHg
               DIA:  10 ~ 220 mmHg     
               MEAN:  20 ~ 235 mmHg 
    Pediatric:  SYS:  40 ~ 235 mmHg
    DIA:  10 ~ 220 mmHg     
    MEAN:  20 ~ 225 mmHg
    Neonate:   SYS:  30 ~ 135 mmHg
    DIA:  10 ~ 110 mmHg     
    MEAN:  20 ~ 125 mmHg
    -Static pressure accuracy: ±3mmHg
    -Resolution: 1mmHg
    -Accuracy: Maximum Mean error ±5mmHg           
             Maximum Standard deviation ≤8mmHg
    -Over pressure Protection: Dual protection via software & hardware 
    Performance Specifications
    Suntech NIBP
    -Measure Method: Oscillometric. Diastolic values correspond to Phase 5 Korotkoff sound
    -Measure Range:
    Adult:   SYS:  40 ~ 260 mmHg
    DIA:  20 ~ 220 mmHg     
    MEAN: 26 ~ 220 mmHg 
    Pediatric:   SYS:  40 ~ 230 mmHg
    DIA:  20~ 160 mmHg     
    MEAN:  26 ~ 183 mmHg
    Neonate:   SYS:  40 ~ 130 mmHg
    DIA:  20 ~ 100 mmHg     
    MEAN:  26 ~ 110 mmHg
    -PR Range: 30 to 220 BPM
    -PR Accuracy: ±2% or ±3 BPM, whichever is greater 
    -Pressure Transducer Accuracy: ±3mmHg between 0 mmHg and 300 mmHg 
    for operating conditions between 0℃ and 50℃
    -Deflate Rate: Deflation step size varies with heart rate, cuff pressure and cuff volume
    -Initial Inflation Pressure: 
    ADULT:  160 mmHg (default) 
    Variable from 120 to 280 mmHg 
    PEDIATRIC:  140 mmHg (default) 
    Variable from 80 to 280 mmHg 
    NEONATE:  90 mmHg (default) 
    Variable from 60 to 140 mmHg
     
     
     
     
     
     
    SpO2 (Digital Technic) 
    -Measurement Range: 0 ~ 100 %
    -Resolution: 1 %
    -Accuracy:  ±2% (70% ~ 100%)
                ±3 % (35% ~ 69%)
    Unspecified (0 ~ 34%)
    -Support Pitch tone and multi-level volume   
    Pulse Rate
    -Measuring and Alarm Range: 25~250bpm
    -Accuracy: ±1% or ±1 bpm, whichever is greater
    -Resolution: 1bpm
     
    Nellcor-SpO2 (Option)
    -Measurement Range: 0 ~ 100 %
    -Resolution: 1 %
    -Accuracy:   70% ~ 100%, ±2 % 